HVI SPECTRUM I
The main control cabinet is 64 cm long and 80 cm deep.
The Length/Strength cabinet is 84 cm long by 80 cm deep.
The Length/Strength cabinet is connected to the main control cabinet by a triangular-shaped support frame that is 70 cm at the widest part.
The printer rests on a shelf located to the right of the main control cabinet that is 50 cm wide and 64 cm deep.
The HVI SPECTRUM I weighs approximately 50 gross kilos (64 net).
Electrical Requirements
The HVI SPECTRUM instrument operates on 208 or 230 volts ± 10% (single phase) (47 to 63 Hz) and requires a separate dedicated 20-amp circuit breaker at the facility’s electrical load center. During normal operation, the HVI SPECTRUM draws approximately eight (8) amps; the startup current is sufficiently high to require the larger breaker. A power cord is provided that will connect to the HVI SPECTRUM. The opposite end must be fitted with connections compatible with the facility’s electrical system. An uninterruptible power supply (UPS) is suggested.
Power Line: Single phase main with protection switch.
If these conditions cannot be met, a power regulator must be installed. It is recommended that you use SOLA #63-23-220-8 (60 Hz), #63-23-720-8 (50 Hz), or an equivalent.
Air Requirements
The system pressure should be regulated between 100psi and 150psi of clean, dry, unregulated compressed air. Use of dryer system, air filters and water trap is highly recommended.
Air Pressure 100 - 120 PSIG (585-825 kPA)
Amount of compressed air approx. 8.5m3/h
Air Tubes Diameter 10 mm (3/8”)
Air Quality Requirements
The air supplied to the instrument should comply with ISO8573.1 Quality Class 3 as follows:
DIRT particle size: < 5 microns
WATER Pressure Dew Point: <-4 F @ 100 PSIG (128 ppm by volume)
OIL (including vapor): < 1.0 ppm